Part Number Hot Search : 
AS5305 N3031 S5771 Z86E73 GM79L05 057524 BYT56A10 LTC36
Product Description
Full Text Search
 

To Download MDS-300ADB48AA Datasheet File

  If you can't view the Datasheet, Please click here to try to view without PDF Reader .  
 
 


  Datasheet File OCR Text:
  medical ac - dc enclosed power supply 48 v 30 0w with 5v/0.5a standby / mds - 300 a db 48 a a technical datasheet all parameters are specified at 25 c ambient unless otherwise noted. www. deltapsu .com ( november 201 6 , rev. 0 4 ) 1 mds - 30 0 a db 48 a a highlights & features C safety approvals to iec 60601 - 1 3 .1 rd ed. & iec 60950 - 1 C compliant with iec 60601 - 1 - 2 4th ed. requirements C up to 21 0 watt convection in 3 .5 " x 5 .5 " x 1. 75 " packag e C up to 800 k hrs mtbf C suited for type bf medical produ cts. C 2 x mopp isolation safety standards cb certified for worldwide use model number: mds - 30 0 a db 48 a a unit weight : 610 g (1.35lb) dimensions ( w x l x h ): 88.9x14 0 . 0 x4 5 . 0 mm ( 3.5x5.5x1.7 5 inch ) general description delta s new mds - 300a d b 48 a a enclosed design offers a high energy density ( 8.8 watts/in 3 ) 3 .5 x5 .5 inch design for type bf patient access medical products. with operating parameters of 90 to 264 vac universal input voltage, temperatures of - 10 to +70 degrees centigrad e, and altitudes of up to 5000 meters (16,400 feet), the design is well suited for a variety of both medical and non - medical applications. other features include input surge of 300 vac, low leakage current, no - load input power < 0.5 watt, and 800k hour mt bf. this product is certified for emc standards en 55011 for industrial, scientific and medical (ism) radio - frequency equipment, and en 55022 for industrial technology equipment (ite) radio - frequency equipment. in addition, only recognized japanese capaci tors are used to ensure long product life. an enhanced feature set, includes a 5 volt/0.5 amp standby output, remote on/off, remote sense, and a power good signal that are included as part of the standard product. the design, which has protection against shock compliant with 2xmopp and type bf requirements, has both medical (with risk report available), and ite safety approvals, including curus(us&canada) /tuv/gb(china); plus, cb certificates with all national deviations. full compliance with rohs direct ive 2011/65/eu for environmental protection is included. model information (all with 5v/0.5a standby available) : model number input v oltage output v oltage conv e ction curr ent output forced air curr ent output mds - 300a d b 48 a a 90 - 2 64 v ac 48 vdc 0 - 4.375 a 0 - 6 .25 a (with 10 cfm forced air) mds 300 a d b 48 a a delta medical power supply max wattage in the product series. 30 0 30 0 w family code a z product type d : en closed input type code b: 3pin class i output voltage 48 for 48 v revision code
medical ac - dc enclosed power supply 48 v 30 0w with 5v/0.5a standby / mds - 300 a db 48 a a technical datasheet all parameters are specified at 25 c ambient unless otherwise noted. www. deltapsu .com ( november 201 6 , rev. 0 4 ) 2 specifications input ratings / characteristics nominal input v oltage 100 - 2 4 0 v ac input v oltage range 90 - 2 64 v ac nominal input f r equency 50 - 60hz input f r equency range 47 - 63hz input cur r ent ( max ) 4 a in put surge voltage (max) 300vac for 10 0 ms full load e f ficiency (typ.) 9 2.5 % @ 115vac /60hz 93.5% @ 230v ac /50hz , reference fig.1 standby power (max) 0.5w (inhibit signal high , 48 volt turned off ) @ 115vac/6 0hz, 230v ac /50hz inrush cur r ent ( max ) 40 a @ 2 3 0 v ac, cold start input - pe(protective earth ) l eakage c ur r ent (max) 0.1ma @ nc 1) , 0. 3 ma @ sfc 1 ) output - pe (protective earth ) l eakage current for type bf application (max) 0.1ma @ nc 1) , 0. 5 ma @ sfc 1 ) power factor (min) 0.9 @ 115v/50hz, 230v/50hz, full load 1) n c: normal condition, sfc: single fault condition leakage current input - pe leakage current 100vac/60hz (typ) 264vac/60hz (typ) delta limit iec60601 - 1 limit normal condition 17.5 ua 43.5ua 100ua max 5000ua max single fault condition 54.5 ua 153 ua 300ua ma x 10000ua max output - pe leakage current for type bf application normal condition 29 .5ua 8 1.5 ua 100ua max 100ua max single fault condition 43.7 ua 129.4 ua 500ua max 500ua max fig.1 efficiency versus output load
medical ac - dc enclosed power supply 48 v 30 0w with 5v/0.5a standby / mds - 300 a db 48 a a technical datasheet all parameters are specified at 25 c ambient unless otherwise noted. www. deltapsu .com ( november 201 6 , rev. 0 4 ) 3 output ratings / characterist ics main output nominal output v oltage (vrated) 48 v total regulation 3 % output power 30 0w 10cfm air, up to 210w convection air line regulation (max) output ratings / characteristics standby output nominal output v oltage of standby output (vrated) 5 v total regulation of standby output 3 % ripple & noise of standby output 100mv max ( refer to fig. 3 )
medical ac - dc enclosed power supply 48 v 30 0w with 5v/0.5a standby / mds - 300 a db 48 a a technical datasheet all parameters are specified at 25 c ambient unless otherwise noted. www. deltapsu .com ( november 201 6 , rev. 0 4 ) 4 fig. 3 ripple & noise measurement circuit mechanical case cover enclosed cover (spcc) dimensions (w x l x h) 88.9x14 0 . 0 x4 5 . 0 mm ( 3.5x5.5x1.7 5 inch ) unit weight 610 g ( 1 . 34 lb) environment surrounding air temperature operating absolute m ax imum/minimum ratin g. - 1 0c to + 7 0c . forced air: l inear ly derate from 100% load at 5 0c , to 50% load at 7 0c (2.5%/ degree centigrade) convection air: linearly derate from 100% load at 5 0c, to 4 5 % load at 7 0c (2. 7 %/ degree centigrade) note: see power de - rating cur ves below storage - 40 c to + 8 5c operating humidity 5 - 9 5 % rh (non - condensing) operating altitude 5 ,000 m eters ( 16,400 feet or 50kpa ) non - o perating altitude 5 ,000 m eters ( 16,400 feet or 50kpa ) shock test (non - operating) 50g, 11ms, 3 shocks for eac h direction vibration (operating) 5 - 500hz, 2grms, 15 minute for each three axis 0 . 4 7 u f c e r a m i c 1 0 u f e l e c t r o l y t i c p s u l o a d t e s t f i x t u r e 5 c m s c o p e
medical ac - dc enclosed power supply 48 v 30 0w with 5v/0.5a standby / mds - 300 a db 48 a a technical datasheet all parameters are specified at 25 c ambient unless otherwise noted. www. deltapsu .com ( november 201 6 , rev. 0 4 ) 5 power de - rating curve s
medical ac - dc enclosed power supply 48 v 30 0w with 5v/0.5a standby / mds - 300 a db 48 a a technical datasheet all parameters are specified at 25 c ambient unless otherwise noted. www. deltapsu .com ( november 201 6 , rev. 0 4 ) 6 protections (f or both 48 volt and 5volt stand b y outputs, unless otherwise indicated) overvoltage (max) main output 1 25 % of rated voltage , lat ch mode standby output 1 25 % of rated voltage , latch mode over load / over current (max) main output 1 6 0 % of rated current standby output 1.06a max hiccup mode ( non - latching , auto - recovery ) over temperature latch mode for main output short circuit hicc up mode, ( non - latching , auto - recovery) reliability data mtbf(minimum) at 115v ac , 2 1 0w, 35 c convection air flow 8 00 kh rs based on telecordia sr - 332 operating life (minimum) at 115vac, 2 1 0w, 25 c convection air flow 26,280 h rs safety standards / directives medical safety iec60601 - 1 2 nd , 3 rd and 3.1 rd ed ition cb report iec60601 - 1 e dition 3.1 rd (2012), en60601 - 1 (2006) + a11 + a1 + a12, can/csa - c22.2 no. 60601 - 1:14, ansi/aami es60601 - 1:2005/(r)2012 ite safety iec60950 - 1 cb report tu v60950 - 1 ul60950 - 1+can/csa60950 - 1 gb4943.1 - 2011, gb9254 - 2008, gb17625.1 - 2003 ce mdd directive 93/42/eec environmental rohs directive 2011/65/eu compliant galvanic isolation input to/ output (2xmopp) input to/ground (1xmopp) output to/ground (1xmopp) 4000 vac 1500 vac 1 ) 1500 vac ( t ype bf application rated) 1) psu can support poe ap plications with primary to fg 2 5 00 vac test.
medical ac - dc enclosed power supply 48 v 30 0w with 5v/0.5a standby / mds - 300 a db 48 a a technical datasheet all parameters are specified at 25 c ambient unless otherwise noted. www. deltapsu .com ( november 201 6 , rev. 0 4 ) 7 emc (compliant with iec - 60601 - 1 - 2 4 th ed. requirements) emc / emissions en550 11/ en550 22, fcc title 47 : class b harmoni c current emissions iec61000 - 3 - 2 meet class d limit immunity to voltage flicker iec 61000 - 3 - 3 electrostatic discharge iec61000 - 4 - 2 level 4 criteria a 1) air discharge: 15 kv contact discharge: 8 kv radiated field iec61000 - 4 - 3 criteria a 1 ) 80mhz - 2700mh z, 10v/ m am modulation 385mhz - 5785mhz, 28v/ m pulse mode and other modulation electrical fast transient / burst iec61000 - 4 - 4 level 3 criteria a 1) : 2kv surge iec61000 - 4 - 5 level 3 criteria a 1) common mode 3 ) : 2kv differential mode 4 ) : 1kv conducted iec6100 0 - 4 - 6 level 2 criteria a 1) 150khz - 80mhz, 3 vrms , 6vrms at ism bands and amateur radio bands power frequency magnetic fields iec61000 - 4 - 8 criteria a 1) magnetic field strength 3 0 a/ m voltage dips iec61000 - 4 - 11 criteria a 1) 0% u t , 0.5 cycle ( 10ms ) , 0 o /45 o /90 o /135 o /180 o /225 o /270 o /315 o /360 o criteria b 2 ) ( criteria a 1) with 1 9 0w load ) 0% u t , 1 cycle ( 20ms ), 0 o criteria b 2) 70 % u t , 25 cycle ( 500ms ), 0 o criteria b 2) 0 % u t , 250 cycle ( 50 0 0ms ), 0 o 1 ) criteria a: normal performance within the specification li mits 2) criteria b : output out of regulation, or shuts down during test. automatically restore to normal operation after test. 3) asymmetrical: common mode (line to earth) 4 ) symmetrical: differential mode (line to line) 5 ) unless otherwise noted, requirem ents apply at 30 0 watt output load
medical ac - dc enclosed power supply 48 v 30 0w with 5v/0.5a standby / mds - 300 a db 48 a a technical datasheet all parameters are specified at 25 c ambient unless otherwise noted. www. deltapsu .com ( november 201 6 , rev. 0 4 ) 8 dimensions mechanical drawing ( mds - 300a d b 48 a a ) notes D dimensions are in mm (inches) D there are two locations where assembled power supply is connected to the customers product a. bottom mounting, use (4x) m3 screws to affix assembled power supply to products enclosure. extruded hole with thread must be with stand 9kgf.cm ( 7 . 81 lb - in) min. maximum allowed screw penetration is 3.5mm (0.138 inch). b. side mounting, use (2x) m3 screws to affix one side of assembled power supply to the products enclosure. extruded hole with thread must be withstand 9kgf.cm ( 7 . 81 lb - i n) min. maximum allowed screw penetration is 3.5mm (0.138 inch).
medical ac - dc enclosed power supply 48 v 30 0w with 5v/0.5a standby / mds - 300 a db 48 a a technical datasheet all parameters are specified at 25 c ambient unless otherwise noted. www. deltapsu .com ( november 201 6 , rev. 0 4 ) 9 input connector cn1 control and standby connector cn101 pin 1 neutral pin 1 remote sense + pin 2 line pin 2 remote sense C cn1 mates with molex housing 26033031 and molex series 6838 crimp terminals. input line can also be connected to pin 1, and input neutral can b e connected to pin 2. pin 3 power good + pin 4 power good - (dc rtn) pin 5 remote on_off/inhibit + pin 6 remote on_off/inhibit - (dc rtn) output connector pin7 5v standby cn102 vo pin 8 dc rtn cn103 dc rtn pin 9 5v standby pin 10 dc rtn cn10 1 mates with molex housing 1041421000 and molex series 104539 crimp terminals. cn2 : pingood jp - 13t mates with kst : fdfnyd1 - 187 or other applicable connectors. s ystem protective earth connection can be connected to the enclosed cover or to cn2.
medical ac - dc enclosed power supply 48 v 30 0w with 5v/0.5a standby / mds - 300 a db 48 a a technical datasheet all parameters are specified at 25 c ambient unless otherwise noted. www. deltapsu .com ( november 201 6 , rev. 0 4 ) 10 functions start - up time the time required for the output vo ltage to reach 90% of its final steady state value, after the input voltage is applied. rise time the time required for the output voltage to c hange from 10% to 90% of its final steady s tate value. hold - up time time between the collapse of the ac input voltage, and the output falling to 9 0 % of its steady state value. graph illustrating the start - up time, rise time, and hold - up time
medical ac - dc enclosed power supply 48 v 30 0w with 5v/0.5a standby / mds - 300 a db 48 a a technical datasheet all parameters are specified at 25 c ambient unless otherwise noted. www. deltapsu .com ( november 201 6 , rev. 0 4 ) 11 dynamic response ( 48 volt output) the power supply output voltage will remain within 5 % of its steady state value, when subjected to a dynamic load 50 to 100% of its rated current. inrush current inrush cu rrent is the input current that occurs when the input voltage is first applied. for ac input voltages, the maximum peak value of inrush current will occur during the first half cycle of t he applied ac voltage. this peak value decreases exponentially during subsequent cycles of ac voltage. overvoltage protection the power supplys overvoltage circuit will be activated when its internal feedback circuit fails. the output voltage shall not exceed its specifications defined on page 5 under protections. power supply will latch off, and require removal/re - application of input ac voltage in order to restart. overload & over current protections the power supplys overload (olp) a nd over current (ocp) protections will be activated before the 48 volt output s current exceeds 1 6 0 % of i o (max load). upon such an occurrence, v o will start to drop. once the power supply has reached its maximum power limit, the protection will be activat ed and the power supply will go into hiccup mode (auto - recovery). the power supply will recover once the fault condition causing the olp and ocp is removed and i o is back within the specified limit. for the standby output, the performance is the same, wi th the exception that the protection will be activated before the output current exceeds 1.06 amps. additionally, if the 48 volts i out is >1 0 0%; but <1 6 0% for a prolong ed period of time (depending on the load), the over temperature protection (otp) wil l be activated due to high temperature on critical components. the power supply will then go into latch mode . short circuit protection the power supplys output olp/ocp function also provides protection against short circuits. when a short circuit is app lied, the output current will operate in hiccup mode, as shown in the illustration in the olp/ocp section on this page. the power supply will return to normal operation after the short circuit is removed . over temperature protection as mentioned above, the power supply also has over temperature protection (otp). this is activated when the overload condition persists for an extended duration and the output current is below the overload trigger point but >100% load. in the event of a higher operating tempe rature condition at 100% load, the power supply will run into otp when the surrounding air temperature is higher than the operating temperature. when activated, the output voltage will go into latch mode until the input voltage is removed; then, reapplied , and the surrounding air temperature drops to its normal operating temperature.
medical ac - dc enclosed power supply 48 v 30 0w with 5v/0.5a standby / mds - 300 a db 48 a a technical datasheet all parameters are specified at 25 c ambient unless otherwise noted. www. deltapsu .com ( november 201 6 , rev. 0 4 ) 12 power good power good+ pin is an open collector transistor (40v/600ma rating). a resistor (suggested value 10kohm, 1/8w) can be added between 5v standby pin (or, oth er available pull - up voltage that is no greater than 30v ) and the power good+ pin (refer to figure below). value of pull - up resistor may have to be adjusted, depending on voltage used, and other end - use conditions of the power good+ pin connection to the p roduct. when ac input is on, power good+ pin will be high. when ac input is off, power good+ pin will be low. there will be a minimum of 5 milliseconds (at 2 1 0w load) between the time the power good goes to low level, and the time when the output reaches 90% of its rated value. remote on_off/inhibit remote on_off/inhibit can be used to enable or disable only the main output. when the main output is disabled, the +5v standby output will continue to operate. this signal can be pulled down to a l ow level of 0.3 volts, or shorted to dc - return, in order for the main output to be enabled; and, floated (no connection to the signal), or pulled up to a value greater than or equal to 3 volts, in order to disable the main output. remote sense remote sense feature can be used to compensate for the extra voltage drop on output wires that are connected from the main output terminals, to the load. with wires connected from the remote sense pins, at the same locations as the wires from the main output, the remo te sense function can compensate up to 500mv voltage drop. the power supply will not be damaged if the remote sense pins are shorted, or if a reverse/inverted polarity connection is made to the load.
medical ac - dc enclosed power supply 48 v 30 0w with 5v/0.5a standby / mds - 300 a db 48 a a technical datasheet all parameters are specified at 25 c ambient unless otherwise noted. www. deltapsu .com ( november 201 6 , rev. 0 4 ) 13 certificate all delta medical power prod ucts conform to the european directive 2011/65/eu. rohs is the abbreviation for restriction of the use of certain hazardous substances delta has been certified as meeting the requirement of iso 13485: 2003 and en iso 13485:2012 for the design and manuf acture of switching power supply and adaptor for medical device. in addition to a ul total certification program (tcp) approved client laboratory for iec60950 and iec60065. delta also has participated ul client test data program (cdtp) for iec 60601


▲Up To Search▲   

 
Price & Availability of MDS-300ADB48AA

All Rights Reserved © IC-ON-LINE 2003 - 2022  

[Add Bookmark] [Contact Us] [Link exchange] [Privacy policy]
Mirror Sites :  [www.datasheet.hk]   [www.maxim4u.com]  [www.ic-on-line.cn] [www.ic-on-line.com] [www.ic-on-line.net] [www.alldatasheet.com.cn] [www.gdcy.com]  [www.gdcy.net]


 . . . . .
  We use cookies to deliver the best possible web experience and assist with our advertising efforts. By continuing to use this site, you consent to the use of cookies. For more information on cookies, please take a look at our Privacy Policy. X